QE65000-ABS Scientific-grade Spectrometer
for Absorbance Measurements

The QE65000-ABS Scientific-grade Spectrometer is a preconfigured combination of detector, optical bench and electronics technologies that provides users with a remarkably sensitive system for demanding low-light level absorbance, reflection, irradiance and transmission measurements.  The QE65000-ABS can achieve up to 90% quantum efficiency with high signal-to-noise and rapid signal processing speed. The QE65000-ABS Scientific-grade Spectrometer is available for $11549.

New! Optical Bench Enhancements

The great performance of our QE65000-ABS Spectrometer has been enhanced to provide even more value: lower stray light, improved efficiency in the UV and Shortwave NIR, and better unit-to-unit reproducibility.

We are also offering a proprietary low BRDF bench coating to improve stray light performance and provide better reproducibility in unit-to-unit performance.

Quantum Efficiency to 90%

The QE65000 is the most sensitive spectrometer we have ever developed. The Hamamatsu FFT-CCD detector used in the QE65000-ABS provides 90% quantum efficiency (defined as how efficiently a photon is converted to a photo-electron). Most of our other detectors are linear CCDs but with this “2D” area detector, we can bin (or sum) a vertical row of pixels, which offers significant improvement in the signal-to-noise ratio (>1000:1) and signal processing speed of the detector compared with a linear CCD, where signals are digitally added by an external circuit.

Increased System Sensitivity

In our spectrometers with linear CCDs, the slit’s width, not its height, regulates the amount of light entering the bench because linear CCDs cannot efficiently collect the light from the entire height of the slit. But in the QE65000, the 2D area detector can better take advantage of the height of the slit and the additional light, greatly improving system sensitivity.

Excellent UV Response

Because the detector in the QE65000-ABS is back-thinned (or back-illuminated), it has great native response in the UV and does not require the additional coatings that we typically apply to other detectors for UV applications.

Demanding Low Light-level Applications

The QE65000-ABS Spectrometer is specially configured for low-light level level absorbance, reflection, irradiance and transmission measurements. The TE-cooled (down to -15 °C) detector features low noise and low dark signal, which enables low-light-level detection and long integration times (analogous to the shutter speed of a camera) from 8 milliseconds to 15 minutes, with virtually no spectral distortion.

Onboard Programming

The QE65000-ABS also has an onboard programmable micro-controller for controlling the spectrometer and accessories. You have access to 10 user-programmable digital inputs/outputs and a pulse generator for triggering other devices. You can use the I/Os to flash a lamp, stop/start a process, and send a message/alarm during the spectrometer’s integration period. The spectrometer’s operating parameters can be controlled through software. In fact, wavelength calibration coefficients unique to each spectrometer are programmed into a memory chip right on the spectrometer.

Plug-and-Play USB Operation

The QE65000-ABS interfaces to a computer via USB 2.0 or RS-232 serial port. Data unique to each spectrometer are programmed into a memory chip on the QE65000; our SpectraSuite Spectroscopy Operating Software reads these values for easy setup and hot swapping among computers, whether they run on Linux, Mac or Windows operating systems.

QE65000-ABS Configuration

The QE65000-ABS is preconfigured with the following, along with SpectraSuite spectrometer operating software and our Annual Service Package (ASP) to extend the warranty from 1 year to 2 years:

Equipment Details
Spectrometer QE65000
Grating HC1-QE (starting at 200 nm)
Filter OFLV-QE-200
Slit SLIT-10

Pricing

Item Description Price
QE65000-ABS QE65000-ABS preconfigured spectrometer system. Includes SpectraSuite and an ASP. $13,528

Specifications

Physical

Dimensions: 182 mm x 110 mm x 47 mm
Weight: 1.18 kg (without power supply)
Detector

Detector:

Hamamatsu S7031-1006

Detector range: 200-1100 nm

Pixels:

1024 x 58 (1044 x 64 total pixels)

Pixel size:

24.576 μm2

Pixel well depth:

1000 Ke-

Sensitivity: ~0.065 counts / e-
Quantum efficiency: 90% peak; 65% at 250 nm
Optical Bench
Design: f/4, Symmetrical crossed Czerny-Turner
Focal length: 101.6 mm input and output
Entrance aperture: 10 µm wide slit
HC1-QE grating: provides 200-950 nm range
OFLV filter: OFLV-QE-200 (200-950 nm)
Collimating and focusing mirrors: Standard only
UV enhanced window: No
Fiber optic connector: SMA 905 to 0.22 numerical aperture single-strand optical fiber
Spectroscopic
Wavelength range: 200-950 nm (Grating dependent)
Optical resolution: ~0.14-7.7 nm FWHM
Signal-to-noise ratio: 1000:1 (at full signal)
A/D resolution: 16 bit
Dark noise: 3 RMS counts
Dynamic range: 7.5 x 109 (system), 25000:1 for a single acquisition
Integration time: 8 ms to 15 minutes
Stray light: <0.08% at 600 nm; 0.4% at 435 nm
Corrected linearity: >99.8%
Electronics
Power consumption: 500 mA @ 5 VDC (no TE cooling); 3.5 A @ 5 VDC (with TE cooling)
Data transfer speed: Full scans to memory every 7 ms with USB 2.0 port, 18 ms with USB1.1 port, 300 ms with serial port
Inputs/Outputs: 10 onboard digital user-programmable GPIOs (general purpose inputs/outputs)
Analog channels: No
Auto nulling: Yes
Breakout box compatibility: Yes, HR4-BREAKOUT
Trigger modes: 4 modes
Strobe functions: No
Gated delay feature: Yes
Connector: 30-pin connector
Power-up time: <5 seconds

Dark current:

4000 e-/pixel/sec @ 25 ºC; 200 e-/pixel/sec @ 0 ºC

Computer
Operating systems: Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port
Computer interfaces: USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 115.2 K baud
Peripheral interfaces: SPI (3-wire); I2C inter-integrated circuit
Temperature and Thermoelectric (TE) Cooling
Temperature limits: 0 ºC to 50.0 ºC; no condensation
Set point: Software controlled; lowest set point is 40 ºC below ambient
Stability +/-0.1 ºC of set temperature in <2 minutes

 

 


PRODUCTS | TECHNICAL | APPLICATIONS | CORPORATE | DISTRIBUTORS | PRIVACY POLICY
CONTACT US
| CATALOG | PRICE LIST | ORDER INFO


.
©2007 Ocean Optics, Inc. All rights reserved.  Terms of Use and Privacy Statement
+1 727.733.2447 • +1 Fax 727.733.3962 •

 

 

 


Contact an
Ocean Optics 
Applications Scientist

 

Convenient Wavelength Calibration! Click here for Details