| Equipment | Details |
| Spectrometer | USB2000+ |
| Grating | 2 |
| Detector | DET2B-200-850 |
| Slit | SLIT-25 |
| Collection Lens | L2 |
| Item | Description | Price |
| USB2000+ | USB2000+RAD preconfigured spectroradiometer. Includes CC3-DA, SPEC-CAL-UV, SpectraSuite and an ASP. | $4678 |
The USB2000+RAD also has an RS-232 serial port for interfacing to PCs and PLCs. Serial port use requires an external power supply and 9-pin serial cable (USB-ADP-PC).
|
Cable |
Description |
Price |
|
USB-CBL-1 |
Cable connects from USB port on USB2000+RAD to USB port on desktop or notebook PC |
$27 |
|
USB-ADP-PC |
Cable and adapter block connect from serial port on USB2000 to serial port on desktop or notebook PC; also comes with USB-CBL-PS 5 VDC power supply |
$104 |
|
CBL-SER |
Spare 9-pin serial cable; connects to USB-ADP-H and serial port on desktop or notebook PC |
$47 |
|
USB-CBL-PS |
Spare 5 VDD power supply; required if USB2000 interfaces to PC via serial port |
$27 |
|
USB-ADP-H |
Adapter block to use with CBL-SER to connect from serial port on USB2000 to serial port on desktop or notebook PC |
$115 |
|
USB-ADP-PX2 |
Adapter block and cable to connect the PX-2 Pulsed Xenon Lamp or LS-450 Pulsed Blue LED to the USB2000+ |
$69 |
| ACC-CON-US2 | Spectrometer accessory connector for external triggering | $0 |
|
Physical |
|
| Dimensions: | 89.1 mm x 63.3 mm x 34.4 mm |
| Weight: | 190 grams |
| Detector | |
|
Detector: |
Sony ILX511 linear silicon CCD array |
| Detector range: | 200-1100 nm |
|
Pixels: |
2048 pixels |
|
Pixel size: |
14 μm x 200 μm |
|
Pixel well depth: |
~62,500 electrons |
| Sensitivity: | 75 photons/count at 400 nm; 41 photons/count at 600 nm |
| Optical Bench | |
| Design: | f/4, Symmetrical crossed Czerny-Turner |
| Focal length: | 42 mm input; 68 mm output |
| Entrance aperture: | 50 µm wide slit |
| Grating: | 2 (250-800 nm) |
| Detector collection lens: | L2 |
| Other bench filter options: | Longpass OF-1 filters |
| Collimating and focusing mirror: | Standard |
| UV enhanced window: | UV2 |
| Fiber optic connector: | SMA 905 to 0.22 numerical aperture single-strand optical fiber |
| Spectroscopic | |
| Wavelength range: | 200 - 850 nm |
| Optical resolution: | ~0.3-10.0 nm FWHM |
| Signal-to-noise ratio: | 250:1 (at full signal) |
| A/D resolution: | 16 bit |
| Dark noise: | 50 RMS counts |
| Dynamic range: | 2 x 10^8 (system); 1300:1 for a single acquisition |
| Integration time: | 1 ms to 65 seconds |
| Stray light: | <0.05% at 600 nm; <0.10% at 435 nm |
| Corrected linearity: | >99.8% |
| Electronics | |
| Power consumption: | 250 mA @ 5 VDC |
| Data transfer speed: | Full scans to memory every 1 ms with USB 2.0 or 1.1 port, 300 ms with serial port |
| Inputs/Outputs: | No |
| Analog channels: | No |
| Auto nulling: | No |
| Breakout box compatibility: | No |
| Trigger modes: | 3 modes |
| Strobe functions: | Yes |
| Gated delay feature: | Yes |
| Connector: | 10-pin connector |
| Computer | |
| Operating systems: | Windows 98/Me/2000/XP, Mac OS X and Linux with USB port; Any 32-bit Windows OS with serial port |
| Computer interfaces: | USB 2.0 @ 480 Mbps; RS-232 (2-wire) @ 57.6 K baud |
| Peripheral interfaces: | I2C inter-integrated circuit |