The new NeoFox Phase Measurement system is a fluorescence-based optical sensor system that helps you reduce costs, improve stability and make calibration easier for a wide variety of oxygen sensing setups. NeoFox is the benchtop device for measurement of fluorescence lifetime, phase and intensity and is particularly suited for applications where sensitivity to drift and system stability are critical.

Plus, NeoFox is about half the cost of comparable phase measurement systems.

NEW!
Check out the FoxyKit sensing kits for probe and patch applications.
 

*Please contact an Applications Scientist before specifying FOSPOR or other coatings.

Probe Specifications:
FOXY Formulation FOSPOR Formulation HIOXY Formulation
  (general-purpose coating) (high-sensitivity for low levels of O2) (for hydrocarbon environments)
O2% range (at 1 ATM): 0-100% 0-<10% 0-20.9%
DO range (ppm at 1 ATM): 0-40 ppm  0-4 ppm 0-8 ppm
Temperature range: -50-+80 ºC for probes 0-+60 ºC for probes -50-+60 ºC
0-60 ºC for patches 0-60 ºC for patches NA
O2% resolution: 100 ppm in gas 10 ppm in gas 100 ppm
DO resolution (at room temp): 4 ppb 0.4 ppb 4 ppb
O2% accuracy: <5% of reading <5% of reading <5% of reading
DO accuracy: <5% of reading <5% of reading <5% of reading
Min. detectable level in gas: 0.01% - 0.05% 0.001% - 0.01% 0.01% - 0.05%
Response time: < 1 second in gas < 30-60 second in gas < 1 second in gas
45-60 seconds with overcoating in gas 60-90 seconds with overcoating in gas  
30-45 seconds in pure water 60-90 seconds in pure water  ~45 seconds in pure water 
Patch Specifications:
  FOXY Formulation FOSPOR Formulation HIOXY Formulation
02% range (at 1 ATM): 0-100% 0-10% 0-20%
DO range (ppm at 1 ATM): 0-40 ppm 0-4 ppm 0-8 ppm
Temperature range: -20 to +60 ºC for patches 0 to +60 ºC for patches -20 to +60 ºC for patches
02% resolution (20 sec averaging): 0.05% 0.01%
(at 30 sec averaging)
0.05%
DO resolution (at room temp): 20 ppb 4 ppb 20 ppb
02% accuracy: 5% of reading 5% of reading 5% of reading
DO accuracy: 5% of reading 5% of reading 5% of reading
Min. detectable level: 0.1% O2 0.01% O2
(at 30 sec averaging)
0.1% O2
Min detectable level in water (at room temp): 40 ppb 4 ppb 40 ppb
Response time: <1 second in gas 30-60 seconds <1 second in gas
  ~30-45 seconds with overcoating in gas ~60-90 seconds with overcoating in gas  
  ~45 seconds in pure water ~60-90seconds in pure water ~30-45 seconds in pure water
Stability (Continuous LED):
  FOXY Formulation FOSPOR Formulation HIOXY Formulation
Lifetime Stability (Tau): 0.0006 usec/ HR 0.003 usec/ HR 0.0002 usec/HR
Oxygen Stability %: 0.01% HR 0.005% HR 0.007% HR
Sterilization Guide:
  FOXY Formulation FOSPOR Formulation HIOXY Formulation
Form factor: Probes/Patches Probes/Patches Probes
Autoclavable: No Not yet determined Yes
Calibration:     Single point reset required after sterilization
Gamma: Yes Not yet determined Yes
Misc:      

 

Get NeoFox Documentation and Software

NeoFox Trouble-shooting Guide


Contact an
Ocean Optics 
Applications Scientist

.