NIRQuest+ Spectrometers
Part Number | Wavelength Range | Optical resolution (FWHM) | Slit Size | Grating Groove Density (lines/mm) | Grating Blaze Wavelength |
---|---|---|---|---|---|
NIRQUEST+1.7-25 | 900 – 1650 nm | 2.8 nm | 25 μm | 150 | 1100 nm |
NIRQUEST+1.7-50 | 900 – 1650 nm | 3.22 nm | 50 μm | 150 | 1100 nm |
NIRQUEST+1.7-100 | 900 – 1650 nm | 5.88 nm | 100 μm | 150 | 1100 nm |
NIRQUEST+1.7-200 | 900 – 1650 nm | 11.06 nm | 200 μm | 150 | 1100 nm |
NIRQUEST+2.2-25 | 900 – 2120 nm | 4.8 nm | 25 μm | 100 | 100 nm |
NIRQUEST+2.2-50 | 900 – 2120 nm | 5.52 nm | 50 μm | 100 | 100 nm |
NIRQUEST+2.2-100 | 900 – 2120 nm | 10.08 nm | 100 μm | 100 | 100 nm |
NIRQUEST+2.2-200 | 900 – 2120 nm | 18.96 nm | 200 μm | 100 | 100 nm |
NIRQUEST+2.5-25 | 900 – 2450 nm | 6.48 nm | 25 μm | 75 | 1700 nm |
NIRQUEST+2.5-50 | 900 – 2450 nm | 7.452 nm | 50 μm | 75 | 1700 nm |
NIRQUEST+2.5-100 | 900 – 2450 nm | 13.608 nm | 100 μm | 75 | 1700 nm |
NIRQUEST+2.5-200 | 900 – 2450 nm | 25.596 nm | 200 μm | 75 | 1700 nm |
Integration time | 1 ms – 120 s | ||||
Connectors | SMA 905 (input fiber) | ||||
Physical dimensions | 182 x 110 x 47 mm | ||||
Weight | 1.15 kg (spectrometer); 450 g (power supply) | ||||
Temperature (operation) | 10 °C – 35 °C | ||||
Temperature (storage) | -30 °C – 70 °C |