Home > Frequently Asked Questions > Products > Accessories > How do the 74-MSP mirror screw plugs increase signal in the cuvette holder for fluorescence measurements?

How do the 74-MSP mirror screw plugs increase signal in the cuvette holder for fluorescence measurements?

It’s not so much that the screw plugs increase signal as they redirect some of the signal that otherwise is lost as it transmits through the sample compartment. Each plug – you’ll need two –screws into a collimating lens port on the cuvette holder to redirect energy back to the sample or back into the collimating lens. The plugs are 0.3″ (7.5 mm) UV-enhanced aluminum-coated reflection mirrors designed to collect the fluorescence that otherwise would be lost and to reflect the excitation energy back through the sample.

CUV-UV-ALL - 74-MSP with CUV-FL-DA