Home > Product Details > USB2000+ Optical Bench Options

USB2000+ Optical Bench Options

USB2000+ Optical Bench Options 1 2 3 4 5 6 7 8 9 10

Hover over the numbers on the image for more info.

SMA 905 Connector
Light from a fiber enters the optical bench.

Click now for more detailed information.

Fixed Entrance Slit
Use our Range and Resolution Calculator to find out how your slit choice affects optical resolution.

Click now for more detailed information.

Longpass Absorbing Filter
A filter has a transmission band and a blocking band to restrict radiation to a certain wavelength region for eliminating second- and third-order effects.

Click now for more detailed information.

Collimating Mirror
Light reflects from this mirror, as a collimated beam, toward the grating.

Click now for more detailed information.

Grating & Wavelength Range
Gratings are fixed in place at the time of spectrometer assembly.

Click now for more detailed information.

Focusing Mirror
This mirror focuses first-order spectra on the detector plane.

Click now for more detailed information.

L2 Detector Collection Lens
This cylindrical lens focuses the light from the tall slit onto the shorter detector elements.

Click now for more detailed information.

Detector
We offer a 2048-element linear CCD array Sony detector.

Click now for more detailed information.

OFLV Variable Longpass Order-sorting Filter
Precisely blocks second- and third-order light from reaching specific detector elements.

Click now for more detailed information.

UV2 Detector Window Upgrade
Standard window is replaced with a quartz window for applications <340 nm.

Click now for more detailed information.

Light from a fiber enters the optical bench through the SMA 905 Connector. The SMA 905 bulkhead provides a precise location for the end of the optical fiber, fixed slit, absorbing filter and fiber clad mode aperture.

Back to Top

Slit PieceLight passes through the installed slit, which acts as the entrance aperture. Slits come in various widths from 5 µm to 200 µm. The slit is fixed in the SMA 905 bulkhead to sit against the end of a fiber. Smaller slit sizes achieve the best optical resolution.

 

 

Slit Description Pixel Resolution
SLIT-5 5-µm wide x 1-mm high ~3.0 pixels
SLIT-10 10-µm wide x 1-mm high ~3.2 pixels
SLIT-25 25-µm wide x 1-mm high ~4.2 pixels
SLIT-50 50-µm wide x 1-mm high ~6.5 pixels
SLIT-100 100-µm wide x 1-mm high ~12 pixels
SLIT-200 200-µm wide x 1-mm high ~24 pixels

Back to Top

If selected, an absorbing filter is installed between the slit and the clad mode aperture in the SMA 905 bulkhead. The filter is used to limit bandwidth of light entering spectrometer or to balance color. Filters are installed permanently.

Item Description
OF1-WG305 Longpass filter; transmits light >305 nm
OF1-GG375 Longpass filter; transmits light >375 nm
OF1-GG395 Longpass filter; transmits light >395 nm
OF1-GG475 Longpass filter; transmits light >475 nm
OF1-OG515 Longpass filter; transmits light >515 nm
OF1-OG550 Longpass filter; transmits light >550 nm
OF1-OG590 Longpass filter: transmits light >590 nm

Back to Top

The collimating mirror is matched to the 0.22 numerical aperture of our optical fiber. Light reflects from this mirror, as a collimated beam, toward the grating. You can opt to install a standard mirror or a UV absorbing SAG+ mirror.

SAG+ mirrors are often specified for fluorescence. These mirrors also absorb nearly all UV light, which reduces the effects of excitation scattering in fluorescence measurements. Unlike typical silver-coated mirrors, the SAG+ mirrors won’t oxidize. They have excellent reflectivity — more than 95% across the VIS-NIR.

Back to Top

GratingsWe install the grating on a platform that we then rotate to select the starting wavelength you have specified. Then we permanently fix the grating in place to eliminate mechanical shifts or drift.

Click here for your Grating & Wavelength Choices.

Predicted Ranges & Resolutions

Use our Range and Resolution Calculator to find out how your grating choice affects spectral range and optical resolution.

Back to Top

This mirror focuses first-order spectra on the detector plane. Both the collimating and focusing mirrors are made in-house to guarantee the highest reflectance and the lowest stray light possible. You can opt to install a standard or SAG+ mirror.

Back to Top

Detector Collection LensThis cylindrical lens is fixed to the detector to focus the light from the tall slit onto the shorter detector elements. It increases light-collection efficiency and reduces stray light. It also is useful in a configuration with a large-diameter fiber for low light-level applications.

Back to Top

We offer a 2048-element (Sony ILX511B) linear CCD array detector with an effective range of 200-1100 nm. Each pixel responds to the wavelength of light that strikes it. Electronics bring the complete spectrum to the software.

Back to Top

Our proprietary filters precisely block second- and third-order light from reaching specific detector elements.

Back to Top

When selected, the detector’s standard BK7 window is replaced with a quartz window to enhance the performance of the spectrometer for applications <340 nm.

Detector Description Spectrometer
DET2B-200-850 Sony ILX511 detector, installed, with 200-850 nm variable longpass filter and UV2 quartz window; best for UV-VIS systems configured with grating #1 or #2 USB2000+
DET2B-200-1100 Sony ILX511 detector, installed, with 200-850 nm variable longpass filter and UV2 quartz window; best for UV-VIS systems configured with XR-1 grating USB2000+
DET2B-350-1000 Sony ILX511 detector, installed, with 350-1000 nm variable longpass filter; best for VIS systems configured with grating #2 or #3 USB2000+
DET2B-UV Sony ILX511 detector, installed, with UV2 quartz window; best for systems for <360nm USB2000+
DET2-VIS Sony ILX511 detector, installed, with VIS BK7 window; best for systems configured for >400nm USB2000+

Detector Specifications

Specification Sony Detector
Detector: Sony ILX511B linear silicon CCD array
Detector range: 200-1100 nm
Pixels: 2048 pixels
Pixel size: 14 µm x 200 µm
Pixel well depth: ~62,500 electrons
Signal-to-noise ratio: 250:1 (at full signal)
Dark noise: 50 RMS counts
Corrected linearity: >99.8%
Sensitivity: 75 photons/count at 400 nm; 41 photons/count at 600 nm