USB-650 Red Tide Spectrometers

$1,220.00

Availability:
View Product Models
Item: USB-650 USB-650-UV USB-650-UV-VIS USB-650-VIS-NIR
Wavelength range: 350-1000 nm 200-880 nm 200-880 nm 350-1000 nm
Cuvette holder-light source: Not included Not included Included Included
Light source: Not applicable Not applicable Deuterium tungsten halogen Tungsten
Couple to optical fibers: Yes Yes No No

Out of stock

SKU: USB-650
Flexible, Economical Choice for Education

USB-650 Red Tide spectrometers are ideal for teaching about UV and visible absorbance, reflectance and emission applications. They are preconfigured, ready-to-use spectrometers that come in several different options.

Model Wavelength Range Light Source Sample Holder Flexibility to Select Accessories
USB-650 350-1000 nm x
USB-650-UV 200-880 nm x
USB-650-UV-VIS 200-880 nm x x
USB-650-VIS-NIR 350-1000 nm x x

  • Smart choice for education – ideal for all kinds of educational lab experiment setups
  • Fully integrated – USB650-UV-VIS and USB650-VIS-NIR models are preconfigured, ready-to-use systems
  • Versatile – USB650 and USB650-UV couple to fibers and other accessories

Physical Specifications

USB-650 Red Tide Spectrometer

Dimensions:

89.1 mm x 63.3 mm x 34.4 mm

Spectrometer weight:

190 g

Detector Specifications

Detector:

Linear silicon CCD array

Pixels:

650 enabled pixels

Pixel size:

14 µm x 200 µm

Pixel well depth:

~62,500 electrons

Sensitivity:

75 photons/count at 400 nm

Optical Bench

Design:

f/4, Asymmetrical crossed Czerny-Turner

Focal length:

42 mm input; 68 mm output

Entrance aperture:

25 µm wide slit

Fiber optic connector:

SMA 905

Spectroscopic

Wavelength range:

USB-650

350-1000 nm

USB-650-UV

200-880 nm

USB-650-UV-VIS

200-880 nm

USB-650-VIS-NIR

350-1000 nm

Optical resolution:

~2.0 nm FWHM

Signal-to-noise ratio:

250:1 (maximum signal)

A/D resolution:

12 bit

Dark noise:

3.2 RMS counts

Integration time:

3 ms to 65 seconds (15 s typical max)

Dynamic range:

8.5 x 107 (system); 1300:1 for a single acquisition

Stray ligth:

<0.05% at 600 nm; 0.10% at 435 nm

Corrected linearity:

>99%